Project No.EN 12543-4:1999
TitleThis European standard specifies the checking of focal spot dimensions above 0,5 mm of X-ray systems up to and including 500 kV tube voltage, by means of radiographs of sharp edges. The image quality and the resolution of X-ray images depend highly on the characteristics of the focal spot. The imaging qualities of a focal spot are based on the two dimensional intensity distribution in the object plane.
Registration number (WIID)4816
ScopeThis European standard specifies the checking of focal spot dimensions above 0,5 mm of X-ray systems up to and including 500 kV tube voltage, by means of radiographs of sharp edges. The image quality and the resolution of X-ray images depend highly on the characteristics of the focal spot. The imaging qualities of a focal spot are based on the two dimensional intensity distribution in the object plane.
StatusIzstrādē
ICS group19.100