<p class="MsoBodyText"><span lang="EN-GB">This document addresses the specification and demonstration of capability and system reliability to successfully deliver a medicinal product as part of design verification and manufacture, utilizing empirical methods as well as modelling. This includes risk-based techniques to establish and demonstrate the required reliability level for all disposable single use drug delivery systems (DDSs) for which there is a single opportunity to deliver a single dose of a medicinal product.</span></p>
<p class="MsoBodyText"><span lang="EN-GB">DDSs covered by this document include, but are not limited to:</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">needle-based injection systems (ISO 11608-1);</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">needle-free injection systems (ISO 21649);</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">aerosol drug delivery systems (ISO 20072).</span></p>
<p class="MsoBodyText"><span lang="EN-GB">This document does not cover:</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">DDSs with containers that can be replaced;</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">DDSs intended for dental use;</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">finished needles;</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">empty syringes;</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">catheters;</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">DDSs for multi-patient use;</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">pumps [IEC 60601-2-24 (if electronic) or ISO 28620 (if non-electronic)];</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">stand-alone prefilled syringes defined by ISO 11040-8.</span></p>
<p class="MsoBodyText"><span lang="EN-GB">In this document system reliability is limited to assessing functional performance of the DDS and does not address human factors or user interface considerations. These will be covered by a usability engineering program in accordance with IEC 62366-1.</span></p>
<p class="Note"><span lang="EN-GB">NOTE<span style="mso-tab-count: 1;"> </span>DDSs not in scope of this document can still benefit from elements in this document but provisions of this document might not completely fulfil the basic safety and effectiveness of such DDSs.</span></p>
Registration number (WIID)
79993
Scope
<p class="MsoBodyText"><span lang="EN-GB">This document addresses the specification and demonstration of capability and system reliability to successfully deliver a medicinal product as part of design verification and manufacture, utilizing empirical methods as well as modelling. This includes risk-based techniques to establish and demonstrate the required reliability level for all disposable single use drug delivery systems (DDSs) for which there is a single opportunity to deliver a single dose of a medicinal product.</span></p>
<p class="MsoBodyText"><span lang="EN-GB">DDSs covered by this document include, but are not limited to:</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">needle-based injection systems (ISO 11608-1);</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">needle-free injection systems (ISO 21649);</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">aerosol drug delivery systems (ISO 20072).</span></p>
<p class="MsoBodyText"><span lang="EN-GB">This document does not cover:</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">DDSs with containers that can be replaced;</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">DDSs intended for dental use;</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">finished needles;</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">empty syringes;</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">catheters;</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">DDSs for multi-patient use;</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">pumps [IEC 60601-2-24 (if electronic) or ISO 28620 (if non-electronic)];</span></p>
<p class="ListContinue1" style="mso-list: l0 level1 lfo1;"><!-- [if !supportLists]--><span lang="EN-GB" style="mso-fareast-font-family: Cambria; mso-bidi-font-family: Cambria;"><span style="mso-list: Ignore;">—<span style="font: 7.0pt 'Times New Roman';"> </span></span></span><!--[endif]--><span lang="EN-GB">stand-alone prefilled syringes defined by ISO 11040-8.</span></p>
<p class="MsoBodyText"><span lang="EN-GB">In this document system reliability is limited to assessing functional performance of the DDS and does not address human factors or user interface considerations. These will be covered by a usability engineering program in accordance with IEC 62366-1.</span></p>
<p class="Note"><span lang="EN-GB">NOTE<span style="mso-tab-count: 1;"> </span>DDSs not in scope of this document can still benefit from elements in this document but provisions of this document might not completely fulfil the basic safety and effectiveness of such DDSs.</span></p>