<p><span lang="EN-GB">This document applies to the basic safety and essential performance of a humidifier, also hereafter referred to as ME equipment, in combination with its accessories, the combination also hereafter referred to as ME system.</span></p>
<p><span lang="EN-GB">This document is also applicable to those accessories intended by their manufacturer to be connected to a humidifier where the characteristics of those accessories can affect the basic safety or essential performance of the humidifier.</span></p>
<p><span lang="EN-GB">EXAMPLE 1 Heated breathing tubes (heated-wire breathing tubes) or ME equipment intended to control these heated breathing tubes (heated breathing tube controllers).</span></p>
<p><span lang="EN-GB">NOTE 2 Heated breathing tubes and their controllers are ME equipment and are subject to the requirements of </span><span lang="EN-GB">IEC</span><span lang="EN-GB"> </span><span lang="EN-GB">60601</span><span lang="EN-GB">‑</span><span lang="EN-GB">1</span><span lang="EN-GB">.</span></p>
<p><span lang="EN-GB">NOTE 3 </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">5367</span><span lang="EN-GB"> specifies other safety and performance requirements for breathing tubes.</span></p>
<p><span lang="EN-GB">This document includes requirements for the different medical uses of humidification, such as invasive ventilation, non-invasive ventilation, nasal high-flow therapy, and obstructive sleep apnoea therapy, as well as humidification therapy for tracheostomy patients.</span></p>
<p><span lang="EN-GB">NOTE 4 A humidifier can be integrated into other equipment. When this is the case, the requirements of the other equipment also apply to the humidifier.</span></p>
<p><span lang="EN-GB">EXAMPLE 2 Heated humidifier incorporated into a critical care ventilator where </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">80601</span><span lang="EN-GB">‑</span><span lang="EN-GB">2-12</span><span lang="EN-GB"> also applies.</span></p>
<p><span lang="EN-GB">EXAMPLE 3 Heated humidifier incorporated into a homecare ventilator for dependent patients where </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">80601‑</span><span lang="EN-GB">2‑72</span><span lang="EN-GB"> also applies.</span></p>
<p><span lang="EN-GB">EXAMPLE 4 Heated humidifier incorporated into sleep apnoea therapy equipment where </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">80601‑</span><span lang="EN-GB">2‑70</span><span lang="EN-GB"> also applies.</span></p>
<p><span lang="EN-GB">EXAMPLE 5 Heated humidifier incorporated into ventilatory support equipment where either ISO 80601-2-79 or ISO 80601-2-80 also apply.</span></p>
<p><span lang="EN-GB">EXAMPLE 6 Heated humidifier incorporated into respiratory high-flow therapy equipment where </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">80601‑</span><span lang="EN-GB">2‑90</span><span lang="EN-GB"> also applies.</span></p>
<p><span lang="EN-GB">This document also includes requirements for an active HME (heat and moisture exchanger), ME equipment which actively adds heat and moisture to increase the humidity level of the gas delivered from the HME to the patient. This document is not applicable to a passive HME, which returns a portion of the expired moisture and heat of the patient to the respiratory tract during inspiration without adding heat or moisture.</span></p>
<p><span lang="EN-GB">NOTE 5 </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">9360</span><span lang="EN-GB">‑</span><span lang="EN-GB">1</span><span lang="EN-GB"> and </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">9360</span><span lang="EN-GB">‑</span><span lang="EN-GB">2</span><span lang="EN-GB"> specify safety and performance requirements for a passive HME.</span></p>
<p><span lang="EN-GB">NOTE 6 </span><span lang="EN-GB">If a </span><span lang="EN-GB">clause</span><span lang="EN-GB"> or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.</span></p>
<p><span lang="EN-GB">Hazards</span><span lang="EN-GB"> inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1.</span></p>
<p><span lang="EN-GB">NOTE 7 Additional information can be found in </span><span lang="EN-GB">IEC</span><span lang="EN-GB"> </span><span lang="EN-GB">60601</span><span lang="EN-GB">‑</span><span lang="EN-GB">1</span><span lang="EN-GB">:</span><span lang="EN-GB">2005</span><span lang="EN-GB">+AMD1:2012+AMD2:2020</span><span lang="EN-GB">, </span><span lang="EN-GB">4.2</span><span lang="EN-GB">.</span></p>
<p><span lang="EN-GB">This document does not specify the requirements for cold pass-over or cold bubble-through humidification devices, the requirements for which are given in ISO 20789.</span></p>
<p><span lang="EN-GB">This document is not applicable to equipment commonly referred to as “room humidifiers” or humidifiers used in heating, ventilation and air conditioning systems, or humidifiers incorporated into infant incubators to humidify the chamber air (i.e., are not directly connected to the patient).</span></p>
<p><span lang="EN-GB">This document is not applicable to nebulizers used for the delivery of a drug to patients.</span></p>
<p><span lang="EN-GB">NOTE 8 </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">27427</span><span lang="EN-GB"> specifies the safety and performance requirements for nebulizers.</span></p>
Registration number (WIID)
87155
Scope
<p><span lang="EN-GB">This document applies to the basic safety and essential performance of a humidifier, also hereafter referred to as ME equipment, in combination with its accessories, the combination also hereafter referred to as ME system.</span></p>
<p><span lang="EN-GB">This document is also applicable to those accessories intended by their manufacturer to be connected to a humidifier where the characteristics of those accessories can affect the basic safety or essential performance of the humidifier.</span></p>
<p><span lang="EN-GB">EXAMPLE 1 Heated breathing tubes (heated-wire breathing tubes) or ME equipment intended to control these heated breathing tubes (heated breathing tube controllers).</span></p>
<p><span lang="EN-GB">NOTE 2 Heated breathing tubes and their controllers are ME equipment and are subject to the requirements of </span><span lang="EN-GB">IEC</span><span lang="EN-GB"> </span><span lang="EN-GB">60601</span><span lang="EN-GB">‑</span><span lang="EN-GB">1</span><span lang="EN-GB">.</span></p>
<p><span lang="EN-GB">NOTE 3 </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">5367</span><span lang="EN-GB"> specifies other safety and performance requirements for breathing tubes.</span></p>
<p><span lang="EN-GB">This document includes requirements for the different medical uses of humidification, such as invasive ventilation, non-invasive ventilation, nasal high-flow therapy, and obstructive sleep apnoea therapy, as well as humidification therapy for tracheostomy patients.</span></p>
<p><span lang="EN-GB">NOTE 4 A humidifier can be integrated into other equipment. When this is the case, the requirements of the other equipment also apply to the humidifier.</span></p>
<p><span lang="EN-GB">EXAMPLE 2 Heated humidifier incorporated into a critical care ventilator where </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">80601</span><span lang="EN-GB">‑</span><span lang="EN-GB">2-12</span><span lang="EN-GB"> also applies.</span></p>
<p><span lang="EN-GB">EXAMPLE 3 Heated humidifier incorporated into a homecare ventilator for dependent patients where </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">80601‑</span><span lang="EN-GB">2‑72</span><span lang="EN-GB"> also applies.</span></p>
<p><span lang="EN-GB">EXAMPLE 4 Heated humidifier incorporated into sleep apnoea therapy equipment where </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">80601‑</span><span lang="EN-GB">2‑70</span><span lang="EN-GB"> also applies.</span></p>
<p><span lang="EN-GB">EXAMPLE 5 Heated humidifier incorporated into ventilatory support equipment where either ISO 80601-2-79 or ISO 80601-2-80 also apply.</span></p>
<p><span lang="EN-GB">EXAMPLE 6 Heated humidifier incorporated into respiratory high-flow therapy equipment where </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">80601‑</span><span lang="EN-GB">2‑90</span><span lang="EN-GB"> also applies.</span></p>
<p><span lang="EN-GB">This document also includes requirements for an active HME (heat and moisture exchanger), ME equipment which actively adds heat and moisture to increase the humidity level of the gas delivered from the HME to the patient. This document is not applicable to a passive HME, which returns a portion of the expired moisture and heat of the patient to the respiratory tract during inspiration without adding heat or moisture.</span></p>
<p><span lang="EN-GB">NOTE 5 </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">9360</span><span lang="EN-GB">‑</span><span lang="EN-GB">1</span><span lang="EN-GB"> and </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">9360</span><span lang="EN-GB">‑</span><span lang="EN-GB">2</span><span lang="EN-GB"> specify safety and performance requirements for a passive HME.</span></p>
<p><span lang="EN-GB">NOTE 6 </span><span lang="EN-GB">If a </span><span lang="EN-GB">clause</span><span lang="EN-GB"> or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.</span></p>
<p><span lang="EN-GB">Hazards</span><span lang="EN-GB"> inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1.</span></p>
<p><span lang="EN-GB">NOTE 7 Additional information can be found in </span><span lang="EN-GB">IEC</span><span lang="EN-GB"> </span><span lang="EN-GB">60601</span><span lang="EN-GB">‑</span><span lang="EN-GB">1</span><span lang="EN-GB">:</span><span lang="EN-GB">2005</span><span lang="EN-GB">+AMD1:2012+AMD2:2020</span><span lang="EN-GB">, </span><span lang="EN-GB">4.2</span><span lang="EN-GB">.</span></p>
<p><span lang="EN-GB">This document does not specify the requirements for cold pass-over or cold bubble-through humidification devices, the requirements for which are given in ISO 20789.</span></p>
<p><span lang="EN-GB">This document is not applicable to equipment commonly referred to as “room humidifiers” or humidifiers used in heating, ventilation and air conditioning systems, or humidifiers incorporated into infant incubators to humidify the chamber air (i.e., are not directly connected to the patient).</span></p>
<p><span lang="EN-GB">This document is not applicable to nebulizers used for the delivery of a drug to patients.</span></p>
<p><span lang="EN-GB">NOTE 8 </span><span lang="EN-GB">ISO</span><span lang="EN-GB"> </span><span lang="EN-GB">27427</span><span lang="EN-GB"> specifies the safety and performance requirements for nebulizers.</span></p>