<p>ISO 13199:2012 specifies the principle, the essential performance criteria and quality assurance/quality control (QA/QC) procedures of an automatic method for measuring total volatile organic compound (TVOC) content in waste gases of stationary sources, using a non-dispersive infrared absorption (NDIR) analyser equipped with a catalytic converter which oxidizes VOC to carbon dioxide.</p>
<p>This method is suitable for the measurement of TVOC emissions from non-combustion processes. This method allows continuous monitoring with permanently installed measuring systems, as well as intermittent measurements of TVOC emissions.</p>
<p>The method has been tested on field operation for painting and printing processes, where TVOC concentrations in the waste gases were from about 70 mg/m<sup>3</sup> to 600 mg/m<sup>3</sup>.</p>
Registration number (WIID)
53494
Scope
<p>ISO 13199:2012 specifies the principle, the essential performance criteria and quality assurance/quality control (QA/QC) procedures of an automatic method for measuring total volatile organic compound (TVOC) content in waste gases of stationary sources, using a non-dispersive infrared absorption (NDIR) analyser equipped with a catalytic converter which oxidizes VOC to carbon dioxide.</p>
<p>This method is suitable for the measurement of TVOC emissions from non-combustion processes. This method allows continuous monitoring with permanently installed measuring systems, as well as intermittent measurements of TVOC emissions.</p>
<p>The method has been tested on field operation for painting and printing processes, where TVOC concentrations in the waste gases were from about 70 mg/m<sup>3</sup> to 600 mg/m<sup>3</sup>.</p>