Project No.ISO 14117:2012
Title<p>ISO 14117:2012 specifies test methodologies for the evaluation of the electromagnetic compatibility (EMC) of active implantable cardiovascular devices that provide one or more therapies for bradycardia, tachycardia and cardiac resynchronization.</p> <p>It specifies performance limits of these devices, which are subject to interactions with EM emitters operating across the EM spectrum in the two following ranges:</p> <p>0 Hz ≤ <i>f</i> &lt; 450 MHz;</p> <p>450 MHz ≤ <i>f</i> ≤ 3 000 MHz.</p> <p>ISO 14117:2012 also specifies requirements for the protection of these devices from EM fields encountered in a therapeutic environment and defines their required accompanying documentation, providing manufacturers of EM emitters with information about their expected level of immunity.</p>
Registration number (WIID)54472
Scope<p>ISO 14117:2012 specifies test methodologies for the evaluation of the electromagnetic compatibility (EMC) of active implantable cardiovascular devices that provide one or more therapies for bradycardia, tachycardia and cardiac resynchronization.</p> <p>It specifies performance limits of these devices, which are subject to interactions with EM emitters operating across the EM spectrum in the two following ranges:</p> <p>0 Hz ≤ <i>f</i> &lt; 450 MHz;</p> <p>450 MHz ≤ <i>f</i> ≤ 3 000 MHz.</p> <p>ISO 14117:2012 also specifies requirements for the protection of these devices from EM fields encountered in a therapeutic environment and defines their required accompanying documentation, providing manufacturers of EM emitters with information about their expected level of immunity.</p>
StatusAtcelts
ICS group11.040.40
33.100.01