Informējam, ka Sistēma pielāgota darbam ar interneta pārlūkprogrammu Internet Explorer (8. un jaunākām versijām) un Mozilla Firefox (3.6 un jaunākām versijām).
Izmantojot citu interneta pārlūkprogrammu, brīdinām, ka Sistēmas funkcionalitāte var tikt traucēta.
This European Standard specifies a method for the determination of decoquinate. This high-performance liquid chromatographic (HPLC) method with a fluorescence detection is applicable to the quantification of decoquinate content in complete and complementary compound feeds, medicated feeds, semi-liquid feeds, premixtures and feed additives.
The method was fully validated from LOQ to 60 000 mg/kg on different matrices during an international collaborative study [11], especially on complete compound feeds for poultry, at trace contamination level of 3 mg/kg and at European authorized level of 20 mg/kg to 40 mg/kg [12].
The limit of detection is between 0,1 mg/kg and 0,3 mg/kg and the limit of quantification is around 0,5 mg/kg. These limits were validated during the collaborative study [11], from results on the blank feed. Lower limits of detection or quantification could be reached but a single laboratory validation is then requested.
Reģistrācijas numurs (WIID)
32274
Darbības sfēra
This European Standard specifies a method for the determination of decoquinate. This high-performance liquid chromatographic (HPLC) method with a fluorescence detection is applicable to the quantification of decoquinate content in complete and complementary compound feeds, medicated feeds, semi-liquid feeds, premixtures and feed additives.
The method was fully validated from LOQ to 60 000 mg/kg on different matrices during an international collaborative study [11], especially on complete compound feeds for poultry, at trace contamination level of 3 mg/kg and at European authorized level of 20 mg/kg to 40 mg/kg [12].
The limit of detection is between 0,1 mg/kg and 0,3 mg/kg and the limit of quantification is around 0,5 mg/kg. These limits were validated during the collaborative study [11], from results on the blank feed. Lower limits of detection or quantification could be reached but a single laboratory validation is then requested.