Informējam, ka Sistēma pielāgota darbam ar interneta pārlūkprogrammu Internet Explorer (8. un jaunākām versijām) un Mozilla Firefox (3.6 un jaunākām versijām).
Izmantojot citu interneta pārlūkprogrammu, brīdinām, ka Sistēmas funkcionalitāte var tikt traucēta.
<p>ISO/TS 13498:2011 establishes a method to determine the torsional yield strength and maximum torque of the implant body/connecting part joints of endosseous dental implant systems. This test is most appropriate for evaluating new types of joints and connecting parts, as well as new materials. </p>
<p>ISO/TS 13498:2011 provides a protocol for torsional loading of an implant body/connecting part joint. It is not applicable for predicting the <i>in vivo</i> performance of an endosseous dental implant system and it is not derived from observations of clinical failure.</p>
Reģistrācijas numurs (WIID)
53823
Darbības sfēra
<p>ISO/TS 13498:2011 establishes a method to determine the torsional yield strength and maximum torque of the implant body/connecting part joints of endosseous dental implant systems. This test is most appropriate for evaluating new types of joints and connecting parts, as well as new materials. </p>