<p>This document specifies procedures for the following:</p>
<p>a) primary particle-sizing calibration for particle sizes 1 µm(c) and larger, sensor resolution and counting performance of liquid automatic particle counters that are capable of analysing bottle samples;</p>
<p>b) secondary particle-sizing calibration using suspensions verified with a primary calibrated APC;</p>
<p>c) establishing acceptable operation and performance limits;</p>
<p>d) verifying particle sensor performance using a test dust;</p>
<p>e) determining coincidence and flow rate limits.</p>
<p>This document is applicable for use with hydraulic fluids, aviation and diesel fuels, engine oil and other petroleum-based fluids. This document is not applicable to particle-sizing calibration using NIST SRM 2806b primary calibration suspensions.</p>
Reģistrācijas numurs (WIID)
72619
Darbības sfēra
<p>This document specifies procedures for the following:</p>
<p>a) primary particle-sizing calibration for particle sizes 1 µm(c) and larger, sensor resolution and counting performance of liquid automatic particle counters that are capable of analysing bottle samples;</p>
<p>b) secondary particle-sizing calibration using suspensions verified with a primary calibrated APC;</p>
<p>c) establishing acceptable operation and performance limits;</p>
<p>d) verifying particle sensor performance using a test dust;</p>
<p>e) determining coincidence and flow rate limits.</p>
<p>This document is applicable for use with hydraulic fluids, aviation and diesel fuels, engine oil and other petroleum-based fluids. This document is not applicable to particle-sizing calibration using NIST SRM 2806b primary calibration suspensions.</p>