Projekta Nr.ISO/PWI 7206-10
Nosaukums<p>This document specifies methods of determining the compressive (fracture) or the tension (disassembly) loads required, under specific laboratory conditions, to cause failure of a modular head system.</p> <p>This document applies to components made of metallic and non-metallic materials, such as femoral heads of partial or total hip-joint replacements of modular construction (i.e. a head/neck conical taper connection).</p> <p>This document excludes methods of examining and reporting the test specimens.</p>
Reģistrācijas numurs (WIID)91363
Darbības sfēra<p>This document specifies methods of determining the compressive (fracture) or the tension (disassembly) loads required, under specific laboratory conditions, to cause failure of a modular head system.</p> <p>This document applies to components made of metallic and non-metallic materials, such as femoral heads of partial or total hip-joint replacements of modular construction (i.e. a head/neck conical taper connection).</p> <p>This document excludes methods of examining and reporting the test specimens.</p>
StatussIzstrādē
ICS grupaNav uzstādīts